Outstanding Performance
Two performance capabilities supporting data reliability: excellent reproducibility from the pump and autosampler, and excellent stability of the column oven and detector.
Easy to Use
Along with simple operation and ease of use, Chromaster offers ease of maintenance, a critical requirement for HPLC.
Robustness
Hitachi's reputation for instrument robustness and reliability continues with the Chromaster, built using stronger materials and manufactured to Hitachi's strict quality control standards.
Pumps: 5110 / 5160
Basic UHPLC models with a wide range of HPLC application. 5110: up to 40 MPa, flow 0.001–9.999 mL/min. 5160: up to 60 MPa, flow 0.001–5.000 mL/min. Up to four mixed solvents with the optional low-pressure gradient unit.
Autosamplers: 5260 / 5280
Both withstand 60 MPa. 5260 uses loop injection (cut, all-volume and full-loop methods) with a built-in washing pump and two-solvent washing, and sample rack control from 1–45 °C with cooling and heating. 5280 uses direct injection, syringe wash delivery, 1–35 °C cooling only. Standard capacity 120 × 1.5 mL and 200 × 1.5 mL respectively.
Column Ovens: 5310 / 6310
Easily accommodates a 300 mm analytical column fitted with a guard column, up to three columns. 5310 horizontal type, 1–85 °C. 6310 vertical type, 4–90 °C, with a built-in three-litre water tank.
Detectors: 5410 to 5610
UV (5410, double-beam, 190–600 nm), UV-Vis (5420, 190–900 nm), Diode Array (5430, 190–900 nm with spectrum acquisition), Fluorescence (5440, Xe lamp, Ex 200–850 nm / Em 250–900 nm), RI (5450, refractive index 1–1.75) and MS (5610).
Supply, Installation & Training
Systems are specified, supplied, installed and commissioned by our resident engineers, with operator training and ongoing method and application support.